Chapter 5: Running Samples Manually 123
5 Under View Reports, select how long to view reports.
6 Under Automatic Saving Options, click the checkbox for each file you want
to save; click the Location button to change the default name or storage
location.
A dialog appears where you can enter a new name or different storage
location. For file types and default names and locations, see page 45.
; Tip For Export Documents, append data by selecting an existing Export
Document file in the Choose Location dialog.
7 Click Accept.
When Data Source = From Data Files, the Test Prefs view appears. Skip
to page 126.
When Data Source = From Cytometer, the FACSComp view appears
showing the name, date, time, and status of the most recent Calib file.
NOTICE These calibration settings are not automatically downloaded to
the cytometer.