124 BD FACSCalibur Instructions for Use
Running BD FACSComp
; Tip If you are running manually, make sure Run With the Loader is deselected
in BD FACSComp preferences.
1 If the current Calib file is old or invalid, click Launch FACSComp.
Otherwise, click Skip FACSComp to proceed to the Samples view where
you can retrieve saved settings and download them to the cytometer
(page 128).
2 At the BD FACSComp Set Up view under Assay Selection, select only the
Lyse/No-Wash (LNW) assay.
3 Prepare BD Calibrite beads and perform instrument QC.
Instrument settings are saved to Calib File.LNW if all parameters pass the
Sensitivity Tests.