EasyManua.ls Logo

CareFusion AVEA - Chapter 5: Volumetric Capnography; Volumetric Capnography

CareFusion AVEA
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AVEA ventilator systems Chapter 4: Monitors, Displays and Maneuvers 119
L2786 Rev. M
Start / Stop
The maneuver shall begin when the START key is actuated. The maneuver shall be
immediately terminated when the STOP key is actuated, a patient effort is detected or the
maneuver tidal volume has been delivered and normal ventilation will resume.
Note:
The initiation of a Pflex Maneuver is delayed by two factors. The total delay is equal to 25% of the breath interval plus
the PEEP Equilibration time set. The PEEP Equilibration time has a range of 030 seconds. If the mandatory breath
rate is 10 bpm, the breath interval is therefore 6 seconds and the delay to initiate Pflex is 1.5 seconds plus the PEEP
equilibration time set. At the default PEEP Equilibration time of 1 second, the total delay is 2.5 seconds in this example.
If the mandatory breath rate is set to 1 bpm, the breath interval is 60 seconds and the delay to initiate Pflex in this
worst-case scenario is 15 seconds, plus the PEEP Equilibration time.
Upper P
flex
and Lower P
flex
determination
Once the maneuver tidal volume has been delivered the ventilator will cycle into exhalation. At the end
of exhalation, the P
AW
/ Vol loop will freeze automatically, the upper and lower inflection points, as well as
the delta P
flex
volume, will be calculated and displayed. The ventilator will return to normal ventilation at
the current ventilator settings.
The user can, should they choose to do so, override the P
flex
values by moving the P
flex
indicators to a
new point along the PV loop and pressing the appropriate set key. The corresponding P
flex
values and
delta P
flex
volume change to represent values based on the current position of the indicators. The
ventilator will store up to four PV loops and their respective inflection points simultaneously.
Note:
Once the values have been redefined by the operator the original values cannot be restored.
Alarms
All currently available alarms shall be active during a P
flex
maneuver except Apnea Interval and I-Time Limit.
To Perform a Pflex Maneuver
The Pflex maneuver allows the clinician to determine opening pressures of the lung during a slow flow volume controlled
breath. Because this maneuver is performed at a slow inspiratory flow rate the effects of respiratory system resistance
are minimized.
Note:
Performance of the Pflex maneuver requires a passive patient. In the event that a patient effort is detected the
ventilator will abort the maneuver and deliver a patient effort detected message while simultaneously returning to
normal ventilation at the current settings.
From the Maneuvers Screen select Pflex
The Pflex maneuver screen allows the operator to set:

Table of Contents

Other manuals for CareFusion AVEA

Related product manuals