156 Chapter 6: Infant Non-invasive Ventilation Operator's manual
L2786 Rev. M
Insp Press
Range: 0 to 30 cmH
2
O
Resolution: 1 cmH
2
O
Default: 5 cmH
2
O
Accuracy: ±2 cmH
2
O
Insp Rise
Range: 1 to 9
(relative control with fast being a setting of 1, and slow a setting of 9)
Default
: 5
Note:
Inspiratory Rise is only available when the Rate is set.
Insp Time
Range: 0.15 to 3.0 seconds
Default
: 0.35 seconds
Rate
Range: Off, 1 to 80 bpm
Default
: Off (nCPAP only)
nCPAP Flow
Flow delivery is under software control and limited to a maximum of 15 LPM.
Advanced Settings
Note:
There are no advanced settings for the primary settings when the Rate is set to Off.
Alarms
Note:
The Alarms Settings window is not available in this mode when the Rate is set to Off. Existing machine alarms and
safety systems will be maintained. During nCPAP support with Rate set to Off, certain alarms will be suspended.