AVEA ventilator systems Chapter 5: Volumetric Capnography 141
L2786 Rev. M
Alarms
Figure 5–7: Capnometry Alarms
High EtCO
2
Creates a low-priority alarm if the monitored EtCO
2
exceeds this setting (see the previous figure).
Range: 6 to 150 mmHg (0.8 20 kPa) or Off
Resolution: 1 mmHg (0.1 kPa)
Default: 60 mmHg (8 kPa)
Note:
The High EtCO
2
alarm must be set at least 5 mmHg (0.7 kPa) above the Low EtCO
2
alarm setting.
Low EtCO
2
Creates a low-priority alarm if the monitored EtCO
2
does not exceed the setting (see the previous figure).
Range: 1 145 mmHg (0.1 19.3 kPa) or Off
Resolution: 1 mmHg (0.1 kPa)
Default: 30 mmHg (4 kPa)
Note:
The Low EtCO
2
alarm must be set at least 5 mmHg (0.7 kPa) below the High EtCO
2
alarm setting.