AVEA ventilator systems Chapter 6: Infant Non-invasive Ventilation 163
L2786 Rev. M
Monitors
In Nasal CPAP, all existing monitors will be suspended, except:
Air Inlet Pressure (Air Inlet)
Oxygen Inlet (O
2
Inlet)
Gas Composition Monitor (F
IO
2
)
The following monitors are only available when the Rate is set to one or greater:
Total Breath Rate
Mandatory Breath Rate
Inspiratory Time
Exhalation Time
I:E Ratio
Peak Inspiratory Pressure
Mean Airway Pressure
PEEP
Nasal CPAP specific monitors
In nCPAP/IMV and when a rate of one or greater is set, the nCPAP level is calculated only over those periods when
control pressure is equal to the nCPAP setting (exhalation, following the end of the pressure decay as determined by
inspiratory rise control).
nCPAP level
Range: 0 to 120 cmH
2
O
Resolution: 1 cmH
2
O
Accuracy: ±3.5% of reading or ±2 cmH
2
O, whichever is greater
CPAP Flow (mean inspiratory flow)
Range: 0–300 LPM
Resolution: 0.1 LPM
Accuracy: ±10%