AVEA ventilator systems Chapter 4: Monitors, Displays and Maneuvers 121
L2786 Rev. M
Note:
If the loop and corresponding data are not saved by the operator, the data will be erased after exiting the maneuver
screen.
AutoPEEP
Maneuver Screen
Figure 4–14: AutoPEEP Maneuver Settings
AutoPEEP is the airway pressure at the end of exhalation immediately prior to the beginning of the next mandatory
inspiration. During the execution of this maneuver the ventilator will execute an expiratory hold in which both the
inspiratory and expiratory valves will be closed. The ventilator will establish the AutoPEEP measurement when the
system pressure reaches equilibration, at the next mandatory breath interval or 5 seconds whichever is shorter.