AVEA ventilator systems Chapter 3: Ventilator Operation 73
L2786 Rev. M
Troubleshooting
Alarm Priority Possible causes Actions
LOW Vte Medium Inspiratory time or flow
inadequate in TCPL
Increase inspiratory time
and/or flow
Inspiratory time too short
due to flow cycling in TCPL.
Increase flow cycle setting
Delivered Pressure has
increased to its upper limit
high pressure limit of 3
cmH
2
O
(±2 cmH
2
O)—due to
changes in ventilator
settings, resistance and/or
compliance.
Increase high pressure limit
or check patient condition