AVEA ventilator systems Chapter 6: Infant Non-invasive Ventilation 157
L2786 Rev. M
Alarms suspended during nCPAP with Rate set to off
Time-based alarms Volume-based alarms Pressure alarms
High Rate
I-Time Limit
I:E Limit
Apnea Interval
High Ve
High Vt
Low Vte
Low Ve
Volume Limit
High Ppeak
Ext High Peak Alarm
Low PEEP
Low Ppeak
Occlusion
Low Ppeak
Range: 1 to 40 cmH
2
O (Neonate setting – nCPAP / IMV)
Default: 7 cmH
2
O
Note:
Not available in nasal CPAP mode unless Rate of one or greater is set.
Alarm threshold setting changes during nCPAP / IMV will not be retained for other ventilation modes following exiting
from nasal CPAP mode.
High Ppeak
Range: 2 to 45 cmH
2
O (Neonate setting – nCPAP / IMV)
Default: 20 cmH
2
O
Note:
Not available in nasal CPAP mode unless a Rate of one or greater is set.
Alarm threshold setting changes during nCPAP / IMV will not be retained for other ventilation modes following exiting
from nasal CPAP mode.