AVEA Troubleshooting
Remove ventilator from patient with any potential problem
Will not pass EST - Fails Leak
Circuit wye not fully occluded
Ensure circuit wye is fully occluded
Check for leaks in circuit and reseat
circuit connections to ventilator.
Replace circuit if necessary.
Filter cartridge not properly seated
Remove exhalation cartridge and
check condition of connections.
Reinstall and recheck. Replace if
Leak in exhalation corner
Replace exhalation diaphragm.
Will not pass EST - Fails O
2
calibration
2
connected properly
Inlet gas pressure too low Verify inlet air and oxygen pressure
above 20psig
2
2
No reading from proximal flow sensor
Sensor / Patient size incompatible
See the operators manual for
correct sensor/mode configurations
Ensure sensor properly connected
at both the patient wye and at the
Loose external connection
Check external connection
Vti > Vte when operating without
proximal flow sensors
Normal Condition when operating
on test lung.
Normal if readings are within
ventilator accuracy specifications
No action required if within
specification
Defective expiratory flow sensor
Clean/replace expiratory flow
sensor
Leak in patient circuit, water
collector or exhalation system
Normal if readings are within
ventilator accuracy specifications
of +/-10%
No action required if within
specification
Defective expiratory flow sensor Clean/replace expiratory flow
sensor
Leak in patient circuit, water
collector or exhalation system
Internal fault Call Technical Service