AVEA ventilator systems 201
L2786 Rev. M
Appendix B: Specifications
Pneumatic Supply
Air or Heliox Supply
Pressure Range: 20 to 80 psig (1.4 to 5.5 bar) (Supply Air)
20 to 80 psig (1.4 to 5.5 bar) (Supply Heliox – 80% / 20% Heliox only)
3 to 10 psig (0.2 to 0.7 bar) (Compressor Air)
Temperature: 5 to 40º
C (41 to 104º
F)
Humidity: Dew Point of gas should be 1.7º
C (3º
F) below the ambient temperature (minimum)
Minimum Flow: 80 L/min at 20 psig (1.4 bar)
Air Inlet Fitting: CGA DISS-type body, No. 1160. NIST fitting per BS-5682:1984 (Air) also available.
Heliox Inlet Fitting: CGA DISS-type body, No. 1180. NIST fitting per BS-5682:1984 (Heliox) also available.
Oxygen Supply
Pressure Range: 20 to 80 psig (1.4 to 5.5 bar) (Supply Oxygen)
Temperature: 5 to 40º
C (41 to 104º
F)
Humidity: Dew Point of gas should be 1.7º
C (3º
F) below the ambient temperature (minimum)
Minimum Flow: 80 L/min at 20 psig (1.4 bar)
Inlet Fitting: CGA DISS-type body, No. 1240. NIST fitting per BS-5682:1984 (O
2
) also available.
Electrical Supply
AC Power Supply
The ventilator operates within specification when connected to the following AC power supplies:
Nominal Voltage Range Frequency Range
100 VAC (85 to 110 VAC) 47 to 65 Hz
120 VAC (102 to 132 VAC) 55 to 65 Hz
230 VAC (196 TO 253 VAC) 47 to 65 Hz
240 VAC (204 TO 264 VAC) 47 to 65 Hz