Recommended Periodic Maintenance
CareFusion is committed to product support. If you have any questions concerning your ventilator’s operation or
maintenance contact your product support representative as shown in Appendix A, Contact Information.
The batteries should undergo discharge/charge cycling quarterly (every three months).
A Preventive Maintenance service should be performed on your AVEA ventilator once per year. Call CareFusion,
Customer Care, at the number given in Appendix A to arrange for a qualified Service Technician to perform this.
WARNING
Electric shock hazard - Do not remove any of the ventilator covers or panels. Refer all servicing to a service
technician authorized by CareFusion.
The annual maintenance will include the following.
Replacement of:
The Air inlet Filter
The Oxygen Inlet Filter
The Compressor Inlet Filter (on compressor equipped models)
The Compressor Outlet Filter (on compressor equipped models)
The Exhalation Diaphragm.
At this time the following maintenance will be performed:
Removal & replacement of the above items
Verification that the following transducers are within calibration specifications:
Air
O2
Blended Gas
Expiratory
Inspiratory
Exhaled Flow delta
Wye flow delta
Auxiliary
Esophageal
Replacement of the O
2
Sensor
Verification Testing to confirm the ventilator is functioning within optimum parameters.
Screen Calibration
Calibration of the Capnostat
Battery Performance Vertification
Every two years the following is recommended to be replaced:
Internal Batteries
External Batteries