AVEA ventilator systems 221
L2786 Rev. M
Appendix F: AVEA Message Bar Text
"Confirm Apnea Settings." Selection of CPAP/PSV or APRV on Mode Select popup when
"Proximal Flow Sensor required."
Acceptance of Volume Limit setting when
Size is NEO, Volume Limit is active, and no Wye Flow Sensor
connected (Varflex or Hotwire).
"Bias Flow insufficient to allow Flow
Acceptance of Bias Flow setting or Flow Trigger setting when
Flow Trigger < (Bias Flow + 0.5 lpm).
"Heliox concentration will change."
Acceptance of % O
2
setting when Heliox is being used.
"Nebulizer not available." Acceptance of Peak Flow setting < 15 lpm when Nebulizer is
active or on pressing of Nebulizer membrane key when Peak Flow
"Confirm inspiratory pressure settings." Selection of Volume Limit control when Volume Limit active (i. e.,
not at default / highest value for patient size).
"Settings restored to defaults."
Patient Accept when New Patient selected.
"Compliance Compensation not active for
NEO."
Size Accept when Size is NEO, and Circ Comp setting is non-
zero.
"Minimum 0.2 sec Inspiratory Time."
Acceptance of any combination of settings that will produce an I-
Time of less than 0.2 seconds.
Acceptance of any combination of settings that will produce an I:E
Ratio of 4:1 or greater.
"Maximum 3 sec Inspiratory Time." Acceptance of any combination of settings when size is NEO that
will produce an I-Time of greater than 3 seconds.
"Maximum 5 sec Inspiratory Time." Acceptance of any combination of settings when size is PED or
ADULT that will produce an I-Time of greater than 5 seconds.
Validation failure, while calibration dialog box is active for selected
"Error saving Serial/Model Number"
Service State Only:
On accept of Serial Number or Model Number Change.
Clear Messages
Validation success, while calibration dialog box is active for
"FCV Characterization in progress."
Service State Only:
On start of Flow Control Valve characterization procedure.
"FCV Characterization complete."
Service State Only:
On successful completion of Flow Control Valve characterization
"FCV Characterization failed."
Unsuccessful completion of Flow Control Valve characterization
procedure. Validation failure characterization and tuning data.
Installed Software Version
Current Time, Date, and Runtime Hours
“DPRAM Comm. Error, Ctrl"
Loss of Communication with Control microprocessor
Print Screen button was pressed; commenced sending screen
data to printer.
"Printer Out of Paper." Print Screen button was pressed, printer reported it is out of
Print Screen button was pressed; printer is not available.
"Printer Error." Print Screen button was pressed; printer reported an error