AVEA ventilator systemsAppendix G: Advanced Pulmonary Mechanics Monitored Parameters 227
L2786 Rev. M
Delta AutoPEEP
AW
(dAutoPEEP
AW
)
Delta AutoPEEP
AW
(dAutoPEEP
AW
), is the difference between airway pressure at the end of an
expiratory hold maneuver and the airway pressure at the start of the next scheduled breath after the
expiratory hold maneuver.
Range: 0 to 50 cmH
2
O
Resolution: 1 cmH
2
O
Note: Requires a passive patient.
Accuracy: ± 2cmH
2
0 or ± 5%, whichever is greater
AutoPEEP
ES
AutoPEEP
ES
is defined as the difference between esophageal pressure measured at the end of
exhalation (PEEP
ES
) minus the esophageal pressure measured at the start of a patient-initiated breath
(P
ES start
) and the sensitivity of the ventilator’s demand system. The sensitivity of the ventilator’s demand
system is the difference between the baseline airway pressure (PEEP
AW
) and the airway pressure when
the patient initiates a breath (P
AW start
).
AutoPEEP
ES
= (PEEP
ES
P
ES start
) (PEEP
AW
P
AW start
)
Range: 0 to 50 cmH
2
O
Resolution: 1 cmH
2
O
Note: Requires an esophageal balloon catheter.
Accuracy: ± 2cmH
2
0 or ± 5%, whichever is greater
Transpulmonary Pressure, Plateau (P
tp
Plat)
The ventilator is capable of calculating and displaying the Transpulmonary pressure during an inspiratory
hold, which is the difference between the airway plateau pressure (P
PLAT AW
) and the corresponding
esophageal pressure.
P
tp
Plat = P
PLAT AW
P
ES
Range: 60 to 120 cmH
2
O
Resolution: 1 cmH
2
O
Accuracy: ± 2cmH
2
0 or ± 5%, whichever is greater
Note:
Requires an inspiratory hold and an esophageal catheter.