AVEA ventilator systems Chapter 2: Unpacking and Setup 31
L2786 Rev. M
Note:
At least one analog ground is required for safe and accurate signal output and input. One analog ground is sufficient
for any and all of the other signals.
Selecting Language .
Touch the language box and use the data dial to select the desired language. Use the Accept key to accept the change.
All text displayed on the LCD screen will automatically be translated to the set language.
Note:
For ease of use all languages appear in their native text in the text selection box on the utilities screen.
Low Vte Alarm Sensitivity
Sets the number of consecutive breaths with an exhaled tidal volume below the Low Vte Alarm setting which are
required to sound the alarm. The default is 3 breaths; the range is 1-5 breaths.
Increase FIO
2
Configures the step increase used during the increase oxygen maneuver. Sets the amount of oxygen the
ventilator will increase above the current set
FIO
2
.
Example: If the Increase
FIO
2
is set at 20%
AND
The set FIO
2
is 40%
WHEN
The increase FIO
2
Maneuver is activated the FIO
2
will increase to 60% for two minutes
after which it will return to 40%.
The default setting for infants is 20% and 79% for Pediatric and Adult applications.
Note:
To achieve 100% delivered FIO
2
during the Increase O
2
maneuver set the Increase FIO
2
setting to its maximum of 79%.
Note:
The settings will be reset to default values when New Patient is selected in the set up menu.