164 Chapter 6: Infant Non-invasive Ventilation Operator's manual
L2786 Rev. M
Graphics
All existing waveforms will be maintained, except for the volume (Vt) waveform. The volume waveform will be selectable
with no functionality, and the loops selection button will be disabled.
Displayed waveforms
• Net Flow (Flow)
Range:
Minimum: -2 to +2 LPM
Maximum: -300 to +300 LPM
Default: -40 to +40 LPM
• Inspiratory Flow / CPAP Flow (Finsp)
Range:
Minimum: -2 to +2 LPM
Maximum: -300 to +300 LPM
Default: -20 to +20 LPM
• Expiratory Flow (Fexp)
Minimum: -2 to +2 LPM
Maximum: -300 to +300 LPM
• Airway Pressure / CPAP Level (Paw)
Minimum: -1 to +2 cmH
2
O
Maximum: -60 to +120 cmH
2
O
Default: -20 to +40 cmH
2
O
• Inspiratory Pressure (PINSP)
Minimum: -1 to +2 cmH
2
O
Maximum: -60 to +120 cmH
2
O