EasyManua.ls Logo

CareFusion AVEA - Adult; Pediatric Adapter View; Pediatric; Neonatal Adapter View

CareFusion AVEA
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
136 Chapter 5: Volumetric Capnography Operator's manual
L2786 Rev. M
6. Insert the airway adapter into the CO
2
sensor. The adapter clicks into place when properly inserted.
Figure 5.5a Adult / Pediatric Adapter
Figure 5.5b Pediatric / Neonatal Adapter
Figure 5–5: Aiarway adaptors
7. Perform the “sensor zero” procedure by following the instructions in the section “Zeroing the CAPNOSTAT 5” on page
144. The zeroing procedure must also be performed when switching between disposable and reusable airway adapters.
8. After the sensor is successfully zeroed, place the airway adapter and sensor into the ventilator circuit between the wye
and endotracheal tube (and any adapters) as shown in the preceding illustration.

Table of Contents

Other manuals for CareFusion AVEA

Related product manuals