44 Chapter 2: Unpacking and Setup Operator's manual
L2786 Rev. M
Adult Setting
Pediatric Setting
Neonate Setting
PSV Tmax
5 sec
0.75 sec
0.35 sec
Machine Volume
(Mach Vol)
0 L
0 ml
0 ml
Volume Limit
(Vol Limit)
2.50 L 500 ml 300.0 ml
Inspiratory Rise
(Insp Rise)
5
5
5
Flow Cycle
0% (off)
0% (off)
0% (off)
T High PSV
Off
Off
N/A
T High Sync
0%
0%
N/A
T Low Sync
0%
0%
N/A
Demand Flow
On
On
On
Alarm Settings
Adult Setting
Pediatric Setting
Neonate Setting
High Rate
75 bpm
75 bpm
75 bpm
High Tidal Volume
(High Vt)
3.00 L
1000 ml
300 ml
Low Tidal Volume
(Low Vt)
0.0 L
0.0 ml
0.0 ml
Low Exhaled
Minute Volume
(Low Ve)
1.0 L
0.5 L
0.5 L
High Exhaled
Minute Volume
(High Ve)
30.0 L/min
30.0 L/min
5.0 L/min
Low Inspiratory
Pressure (Low
Ppeak)
8 cmH
2
O 8 cmH
2
O 5 cmH
2
O
High Inspiratory
Pressure (High
Ppeak)
40 cmH
2
O 40 cmH
2
O 30 cmH
2
O
Low PEEP
3 cmH
2
O 3 cmH
2
O 1 cmH
2
O
Apnea Interval
20 sec
20 sec
20 sec
Auxiliary Controls
Adult Setting
Pediatric Setting
Neonate Setting
Manual Breath
---
---
---
Suction
---
---
---
↑ O2
79%
79%
20%
Nebulizer
---
Inspiratory Hold
(Insp Hold)
---
---
---
Expiratory Hold
(Exp Hold)
---
---
---