204 Appendix B: Specifications Operator's manual
L2786 Rev. M
Scale Factor: 1 L/min / 25 mV (Adult)
1 L/min / 50 mV (Pediatric)
1 L/min / 100 mV (Neonate)
Accuracy: ± 10% of reading or ± 30 mV, whichever is greater
Zero Offset: None
3. Volume:
Range: -1.00 to 4.00 L (Adult)
-200 to 800 ml (Pediatric)
-100 to 400 ml (Neonate)
Scale Factor: 1 L / V (Adult)
1 ml / 5 mV (Pediatric)
1 ml / 10 mV (Neonate)
Accuracy: ± 10% of reading or ± 30 mV, whichever is greater
Zero Offset: 1.000 VDC
4. Breath Phase
The ventilator provides a continuous analog voltage representative of breath phase (Inspiration = 5 VDC, Expiration
= 0 VDC).