AVEA ventilator systems Chapter 2: Unpacking and Setup 43
L2786 Rev. M
Default Settings for Adult, Pediatric and Neonate
The Default settings are the operational settings that take effect when you press the New Patient button on power up.
Ventilation Setup
Adult Setting
Pediatric Setting
Neonate Setting
ET tube Diameter
7.5 mm
5.5 mm
3.0 mm
ET Tube Length
30 cm
26 cm
15 cm
Artificial Airway
Compensation
Off Off Off
Leak
Compensation
Off
Off
Off
Circuit
Compliance
Compensation
(Circ Comp)
0.0 ml/cmH
2
O 0.0 ml/cmH
2
O 0.0 ml/cmH
2
O
NOT active in
Neonates.
Humidification
Active On
Active On
Active On
Patient Weight
1 kg
1 kg
1 kg
Primary Controls
Adult Setting
Pediatric Setting
Neonate Setting
Breath Type/Mode
Volume A/C
Volume A/C
TCPL A/C
Breath Rate (Rate)
12 bpm
12 bpm
20 bpm
Tidal Volume
(Volume)
500 ml 100 ml 2.0 ml
Peak Flow
60 L/min
20 L/min
8 L/min
Inspiratory
Pressure (Insp
Pres)
15 cmH
2
O 15 cmH
2
O 15 cmH
2
O
Inspiratory Pause
(Insp Pause)
0.0 sec
0.0 sec
0.0 sec
Inspiratory Time
(Insp Time)
1.0 sec 0.75 sec 0.35 sec
PSV
0 cmH
2
O 0 cmH
2
O 0 cmH
2
O
PEEP
6 cmH
2
O 6 cmH
2
O 3 cmH
2
O
Inspiratory Flow
Trigger (Flow
Trig)
1.0 L/min
1.0 L/min
0.5 L/min
%O
2
40%
40%
40%
Advanced Settings
Adult Setting
Pediatric Setting
Neonate Setting
Vsync
0 (off)
0 (off)
N/A
Vsync Rise
5
5
N/A
Sigh
0 (off)
0 (off)
N/A
Waveform
1 (Dec)
1 (Dec)
1 (Dec)
Bias Flow
2.0 L/min
2.0 L/min
2.0 L/min
Inspiratory
Pressure Trigger
(Pres Trig)
3.0 cmH
2
O 3.0 cmH
2
O 3.0 cmH
2
O
PSV Rise
5
5
5
PSV Cycle
25%
25%
10%