EasyManua.ls Logo

CareFusion AVEA - Page 173

CareFusion AVEA
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AVEA ventilator systems Chapter 6: Infant Non-invasive Ventilation 159
L2786 Rev. M
High nCPAP pressure
A high priority audible/visual alarm will be activated whenever the nCPAP exceeds the high airway pressure threshold for
a period greater than 15 seconds.
Threshold: Set CPAP Level + 3 cmH
2
O
Tolerance: ± 0.5 cmH
2
O
Note:
Threshold is automatically updated upon acceptance of a change in the nCPAP control setting.
Low nCPAP Pressure
A high priority audible/visual alarm will be activated whenever the nCPAP pressure falls below the low airway pressure
threshold for a period greater than 15 seconds.
Threshold: Set nCPAP Level - 2 cmH
2
O (if set nCPAP ≥ 3 cmH
2
O)
Set nCPAP Level - 1 cmH
2
O (if set nCPAP < 3 cmH
2
O)
Tolerance: ± 0.5 cmH
2
O
Note:
Threshold is automatically updated upon acceptance of a change in the nCPAP control setting.
nCPAP Pressure Limit
A high priority audible/visual alarm will be activated whenever the nCPAP pressure is greater than airway pressure limit
for 3 seconds. The alarm will deactivate when the nCPAP pressure drops below 4.5 cmH
2
O.
Pressure Limit: 11 cmH
2
O (nCPAP only, rate set to off)
Set CPAP Level + Inspiratory Pressure + 3 cmH
2
O (nCPAP / IMV, rate not zero)
Tolerance: + 0.5 cmH
2
O

Table of Contents

Other manuals for CareFusion AVEA

Related product manuals