180 Chapter 7: Alarms and Indicators Operator's manual
L2786 Rev. M
becomes disconnected from the slave ventilator during ILV.
INVALID GAS ID A medium priority audible/visual alarm shall be activated, and
INVALID GAS I.D. shall be indicated whenever a defective gas
I.D. connector is installed in the ventilator. When a defective Gas
I.D. connector is detected, the gas corrections default to air.
N/A Medium
FAN FAILURE A low priority audible/visual alarm is activated, and FAN FAILURE
indicated, whenever the fan has stopped rotating.
N/A Low
High EtCO
2
Low priority alarm if monitored EtCO
2
exceeds the alarm
threshold setting.
6 – 150 mmHg or off
Default: 60 mmHg
Low
Low EtCO
2
Low priority alarm if monitored EtCO2 does not exceed the alarm
threshold setting.
1 – 145 mmHg or off
Low