196 Appendix A: Contact and Ordering Information Operator's manual
L2786 Rev. M
3. Carefully ease the red fuse holder out of the power entry module.
4. The fuse holder contains two fuses, 3.1 amps (for 100/120 volt lines) and 2.0 amps (for 230/240 volt lines) as shown in
Table 8–1.
5. Replace the failed fuse in the fuse holder with a fuse whose type, voltage rating, and current rating is identical to the
fuses supplied from the factory.
6. Carefully replace the red fuse holder into the power entry module. Check to ensure that the correct line voltage is
uppermost as you re-insert the fuse holder into the power entry module.
7. Close the power entry module cover and check to make sure that the correct voltage is displayed through the window.