48 Chapter 2: Unpacking and Setup Operator's manual
L2786 Rev. M
Improper charge level indicator -
External battery
Excessively discharged battery
Requires at least 12 hours for full
charge
Decreased run time on battery
Battery not fully charged
Internal battery requires at least 4
hours for full charge. External
battery requires a minimum of 12
hours for a full charge.
Does not run on compressor
Improper sensitivity settings
Check flow and pressure trigger
settings
Perform EST and correct leaks as
required. Bias Flow should be set to
approximately 1.5 lm greater than
Low gas alarm on compressor
Minute volume exceeds 40 L/min
Air/Heliox connector not properly
connected
Device Error indicator Defective sensor Replace sensor
Exhalation flow sensor not
connected
2
connected
2
Defective O
2
sensor Replace O
2
sensor
Improper connection sequence External battery connection should
be made with AC power
Occlusion of expiratory limb of
patient circuit.
Occluded expiratory filter
Check expiratory limb for kinks
and/or water
Replace expiratory filter