Setup
230 FTBx-88000 Series
BERT and Unframed BERT - CPRI/OBSAI
Bit Error
Pass/Fail Verdict enables and sets the bit error rate/count threshold
prior to run the test. This allows for a simple pass/fail verdict, leaving
no room for misinterpretation of the test results. To enable the pass/fail
verdict, select either Bit Error Count or Bit Error Rate (default is
Disabled).
BER Threshold allows entering the maximum bit error count/rate
allowed before declaring a fail verdict: 0 (default) to 999999 for count;
1.0E-14 to 1.9E-01 (default is 1.0E-12) for rate.
CPRI/OBSAI
Note: Available with Framed L2.
Pass/Fail Verdict check box, when selected (default), enables the pass/fail
verdict. A pass verdict is declared when the link status is up and no
CPRI/OBSAI alarm and/or error are recorded for the duration of the test. For
CPRI 24.3G with RS-FEC enabled, the FEC-COR-CW and Pre-FEC-SYMB
errors are not considered in the pass/fail verdict since they are not causing
any traffic error.