Setup
Power Blazer 321
Labels
Labels
For SONET/SDH BERT and SONET/SDH - DSn/PDH BERT: From the test
menu, tap Setup, Test Configurator, the interface block, and on the
Labels tab.
For OTN-SONET/SDH BERT: From the test menu, tap Setup, Test
Configurator, the protocol block, and on the Labels tab.
Labels
Note: Selecting a Label byte to be generated automatically update the
corresponding OH byte and vice versa. Refer to OH - SONET/SDH on
page 710 for more information. Check boxes are coupled with settings from
Labels on page 491.
STS Path (C2) / AU Path (C2): The C2 byte is allocated to indicate the
content of the STS SPE / VC, including the status of the mapped
payloads.
Generated allows the selection of C2 byte (refer to C2 on page 718).
PLM-P/UNEQ-P / HP-PLM/HP-UNEQ check box when selected
(cleared by default) enables the Payload Mismatch and STS/AU UNEQ
monitoring.
Expected allows the selection of the expected C2 byte value (refer to
C2 on page 718).