Setup
Power Blazer 239
BERT - FlexE BERT
Bit Error
Pass/Fail Verdict enables and sets the bit error rate/count threshold
prior to run the test. This allows for a simple pass/fail verdict, leaving
no room for misinterpretation of the test results. To enable the pass/fail
verdict, select either Bit Error Count or Bit Error Rate (default is
Disabled).
BER Threshold allows entering the maximum bit error count/rate
allowed before declaring a fail verdict: 0 (default) to 999999 for count;
1.0E-14 to 1.9E-01 (default is 1.0E-12) for rate.