EasyManua.ls Logo

Fukuda DS-8500 - FLOW-i Data Input

Default Icon
230 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-13
Chapter 4 Connection to the External Devices FLOW-i Data Input
External Device Setup
To display the INVOS 5100C data, "External Device" setup is required.
REFERENCE
Refer to the following procedure. (
"External Device Setup" P4-16)
FLOW-i Data Input
By connecting the FLOW-i Anesthesia Delivery System (MAQUET) to the DS-8500 System, the numeric data,
waveform data of the FLOW-i can be monitored.
The software version of FLOW-i that can be connected to the DS-8500 System is as follows.
System Software Version 02 (FCI Protocol Version 0004)
System Software Version 03 (FCI Protocol Version 0005)
WARNING
The alarm data of FLOW-i cannot be monitored.
The alarm will not be generated on the DS-8500 System.
CAUTION
The operation of anesthesia delivery system should be performed by well-trained and
authorized personnel.
When connecting the DS-8500 System and the anesthesia delivery system, use only the
specified connection cable.
Make sure that the anesthesia delivery system is connected to the specified connector on
the DS-8500 System.
When connecting the cable, make sure that the main power of this equipment and the
anesthesia delivery system is turned OFF.
Connection with the FLOW-i
CAUTION
Only one anesthesia system can be connected to each DS-8500 system. Do not connect
more than one anesthesia systems.
The anesthesia system and Multigas Unit (MGU-800, MGU-810 series) cannot be
connected simultaneously.
The anesthesia system and ventilator cannot be connected simultaneously.
Anesthesia Delivery System
Connection Cable
For Status II Connector For Serial Connector
FLOW-i CJ-402RI-70SVi (x1) CJ-502 (x1)

Table of Contents