Chapter 8 Cleaning/Disinfecting/Storing Cleaning the Equipment and Sensors
BISx
Clean any spillage of blood or solutions on BISx as soon as possible.
Use lint-free absorbent towels for spill cleanups.
Dampen the towel with detergent and lukewarm water to aid in cleaning.
After cleaning, wipe the PIC connector ends with alcohol and allow to dry completely.
Residual moisture inside the connector may affect BISx performance.
Use lint-free absorbent towels dampened with a 10% bleach solution, or a commercial disinfectant.
BIS sensor is disposable. Do not reuse it.
Do not reuse the BIS sensor to other patients. It may cause cross-infection.
The duration for one usage should be within 24 hours.