Chapter 1 Installation of the Unit Precautions for Installing the Equipment
Chapter 1 Installation of the Unit
Precautions for Installing the Equipment
This section describes the environmental condition to use this equipment.
CAUTION
The installation of this equipment should be performed by our service representative or a
person who is well acquainted with this equipment.
Avoid stacking multiple equipments or placing anything on the equipment during operation.
Install this equipment in a place where power supply cable can be easily disconnected.
If any abnormality is found on the equipment, immediately turn OFF the power, and
disconnect the power supply cable from the outlet.
Do not soak or immerse the equipment in liquids.
Operating Environment
The following environmental conditions should be observed when using the equipment.
Surrounding Temperature: 10°C to 40°C (10°C to 35°C for MGU-800/810 series)
Relative Humidity: 30% to 85% (non-condensing)
Atmospheric Pressure: 70 kPa to 106 kPa
This equipment is intended for patient monitoring in NICU, ICU, CCU, surgery, emergency room and ward.
Do not use in MRI environment or in a home-care setting.
The power source should fulfill the following condition.
Use a hospital grade outlet (3-pin grounded outlet).
Verify power voltage and frequency before connecting to an AC power source.
Use the power source that can provide adequate power to the equipment.
Refer
Operation Manual "Specification" P14-1 for power voltage, frequency, and power consumption.
Pay attention to install or store the equipment in proper location. Do not install or store in the following
locations.
where chemicals are stored or gas may generate
where the equipment will be subject to splashing water or humidity from a nebulizer or vaporizer
where the equipment will be subject to direct sunlight
Unstable place with inclination, vibration, or shock
Ensure proper ventilation to cool the equipment.
A minimum space of 5 cm is required between vents on the rear side of the main unit and the wall. If the
main unit is embedded in a wall or surrounded by a wall, a minimum space of 10 cm is required.
The Super Unit (HS-8000 series), DS-8007 system, Multigas Unit (MGU-800/MGU-810 series), Recorder
Unit (HR-800), Input Box (IB-8004) must be transversely installed.If installed in incorrect direction, water or