6-1
Chapter 6 Setup Item/Default Value Setup Item
Chapter 6 Setup Item/Default Value
Setup Item
This section lists selection, default setting, and backup status for each setup item.
The following indicates the selection, default setting and backup status for each setup item.
Initial Settings
Initial Settings (Alarm)
Item Description Default Backup
Alarm System Fukuda Tone, Melodic Tone,
Standard Tone
Standard Tone
○
Basic Alarm Parameter Each Parameter (S, D, M can be
specified for BP)
HR, SpO
2
, NIBP-S,
EtCO
2
○
Asystole, VF, VT Alarm Always ON, ON/OFF Always ON ○
HR/PR Lower Limit during Alarm Auto Setting OFF, 30 bpm, 40 bpm OFF ○
Buzzer Tone at Speaker Failure Enable, Disable Enable ○
Suspend Arrhy. Analysis during Noise Interference ON, OFF OFF ○
Oxygenator
Mode
Alarm All OFF, Sel. Parameter, No Change All OFF ○
Oxygenator Mode Message ON, OFF ON ○
Numeric Data BP1 to BP8, T1 to T8, Tb ○
MODE: Sub Mode OFF, INDUCT., SURGERY,
WAKING, 12LEAD, SUB5, SUB6
OFF ○
Mode: Extended Display 1 OFF, MACHINE, STAFF, EXT1-3 OFF ○
Mode: Extended Display 2 OFF, CIRCUL., ECG CAS., EXT2-3 OFF ○
Alarm Sound: Check Electrodes Enable, Disable Enable ○
Alarm Sound: Check SpO
2
Sensor Attach.
Enable, Disable Enable ○
Lower Limit for
Alarm Volume
Vital Alarm: Urgent 11 levels 1
○
Vital Alarm: Caution 1
Vital Alarm: Status 1
Ventilator Alarm 1
Status Alarm: Urgent 1
Status Alarm: Caution 1
Status Alarm: Status 1
Other Bed Alarm 1
Alarm Indicator
Setup
Level S
*1
Pattern A to J, OFF Pattern A ○
Level H Pattern A ○
Level M Pattern E ○
Level L Pattern F ○
Ventilator Alarm Pattern A ○
Synchronize with HR/RR Sync. to HR, Sync. to RR, OFF OFF ○