xxv
Safety Electromagnetic Compatibility
Electromagnetic Compatibility
The performance of this equipment under electromagnetic environment complies with IEC 60601-1-2:2007.
Do not use any unauthorized equipment or cables as they may not comply with the EMC standard.
Precautions for Safe Operation under Electromagnetic Influence
If any sorts of electromagnetic wave, magnetic field, or static electricity exist around the device, noise interference
or malfunction of the device may occur. If any unintended malfunction or noise occurs during monitoring, check the
electromagnetic influence and take appropriate countermeasures.
The following are examples of the common cause and countermeasures.
Cellular Phone
The radio wave may cause malfunction to the device.
Cellular phones and radio sets should be turned off in the room (building) where medical device is located.
Lightning
A lightning nearby may induce excessive voltage to the equipment. If any danger is suspected;
Use the uninterruptible power supply system.
High frequency noise interference from other device through the power outlet
Check where the noise is originated and remove it using filtering device, etc.
Stop using the device that is originating the noise.
Use other power outlet.
CAUTION
WARNING
WARNING
CAUTION