8-3
Chapter 8 Cleaning/Disinfecting/Storing Cleaning the Equipment and Sensors
3Press again the [Key Lock] key for more than 2 seconds.
The message will disappear, and the key locked condition will be canceled.
Housing
Cleaning
Wipe using a tightly squeezed cloth saturated with diluted neutral detergent. Then wipe with a dry cloth.
Disinfection
Wipe with a cloth dampened with one of the following chemicals. Then, wipe off with dry cloth.
Clean the equipment frequently so stains can be removed easily.
To prevent injury, it is recommended to wear gloves when cleaning the equipment.
When cleaning or disinfecting, do not allow chemical solution to enter the equipment or
connectors.
Do not use organic solvents, thinner, toluene or benzene to avoid damaging the resin case.
Do not polish the equipment with abrasive, chemical cleaner, alkaline or acidic detergent.
Otherwise, the surface resin or paint coating may be damaged, resulting in discoloration,
scratches, and other problems.
For the precautions for storing and handling the chemicals, refer to the instruction manual
for the respective chemical.
Usable Cloth:
*Soft cloth (cotton)
*Soft non-woven cloth (pulp, rayon, polyethylene, etc.)
Chemicals:
*Glutaral 2%
*Alcohol (ethanol, isopropyl alcohol for disinfection)
*Benzalkonium Chloride 0.2%
*Benzethonium Chloride 0.2%
*Alkyldiaminoethylglycine Hydrochloride 0.5%
Usable Cloth:
*Soft cloth (cotton)
*Soft non-woven cloth (pulp, rayon, polyethylene, etc.)