EasyManua.ls Logo

Fukuda DS-8500 - Analog Output Setup

Default Icon
230 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-22
Chapter 4 Connection to the External Devices Setup for the External Device Connection
Analog Output Setup
The HS-8000, DS-8007 is capable to output the analog ECG and BP waveform.
The BP waveform for analog output can be selected from the measured waveforms on the HS-8000 multiparameter
connector.
On the "Analog Output" setup menu, initial settings for display/printing can be performed.
When the HS-8000 is used
1Press the [Menu], [Initial Settings], [External Device], [Analog Output] keys.
The "Analog Output" setup menu will
be displayed.
2Set the ECG waveform lead.
1 Select from [Disp. Lead]/[Selected
Lead].
2 When [Selected Lead] is selected,
press the key for "Output Lead Sel."
to select the output lead.
Select from [I]/[II]/[III]/[avR]/[aVL]/
[aVF]/[V1]/[V2]/[V3]/[V4]/[V5]/[V6].
3Select the IBP waveform to output from the HS-8000.
Select from [MPA1-1] / [MPA1-2] / [MPA2-1] / [MPA2-2] / [MPA3-1] / [MPA3-2].
4Set the synchronized signal output.
1 Press the key for "Signal Output", and select from [HR]/[RR].
[HR]: Synchronized signal based on HR source (ECG, BP, SpO
2
) will be output.
[RR]: Synchronized signal based on RR source (impedance) will be output.
2 Set the "Output Logic".
[Positive Logic]: Positive synchronized signal will be output.
[Negative Logic]: Negative synchronized signal will be output.
REFERENCE
Refer to the operation manual of the HS-8000 Super Unit for connector pin
assignments of the output signal.
2
3
4

Table of Contents