9-11
Chapter 9 Maintenance Check Program Version
When EtCO
2
measurement is not stable or accuracy is degraded compared with other
measuring device.
When the patient monitor was not used for a while, or when EtCO
2
was not measured
for a while.
When "HCP-800 Calibration" is displayed on the "Periodic Replacement Parts" screen
at power ON.
Dispose of calibration gas according to the regulation of each medical institution.
Program Version
On the program version screen, software version of the main unit and modules can be verified.
1Press the [Menu], [Maintenance] keys.
The software version screen will be displayed.
The software version, boot version, date, comment required for the DS-8500 system will be displayed.
DSC-8500 Main Unit Software
Display Unit (LC-8019T/LC-8019TC, LC-8015T/LC-8015TC)
HS-8000/DS-8007 Software
HR-800 Recorder Unit Software
MGU-800/810 Multigas Unit Module, Spiro Unit Module Software
[IB1], [IB2]: The information of the IB-8004 and the modules connected to the IB-8004 will be displayed.
[Serial]: The information of the equipment connected to the serial connector of the main unit will be
displayed.