8-5
Chapter 8 Cleaning/Disinfecting/Storing Cleaning the Equipment and Sensors
it before applying it to the patient. For details of the cleaning procedure, refer to the instruction manual of the
sensor.
Disinfect the Masimo reusable sensor and patient cable according to the manufacturer's guidelines.
When cleaning the Masimo reusable sensor and patient cable, disconnect them from the main unit, and follow
the procedure below.
1Wipe the sensor and cable using 70% isopropyl alcohol cotton.
2Dry it completely with air before reusing.
Temperature Probe
Disinfect the temperature probe according to the manufacturer's guidelines.
When cleaning the relay cable, follow the procedure below.
1Wipe the cable using 70% isopropyl alcohol cotton.
2Dry it completely with air before reusing.
Cardiac Output Relay Cable
Do not reuse / resterilize the cardiac output catheter.
When cleaning, follow the procedure below.
1Wipe the cable using 70% isopropyl alcohol cotton.
2Dry it completely with air before reusing.
Airway Adapter for Capnostat 5
Sterilize the airway adapter according to the manufacturer’s guidelines.
Do not reuse / re-sterilize the disposable airway adapter.
Water Trap (Multigas Unit)
The water trap on the MGU-800/810 series receives fluids from the sampling line connected to the patient.
When the sampling line or water trap gets completely occluded with water, <GAS Check Sample Line> will
be displayed.
No parts of the water trap are intended to be cleaned.
Under environment of 37 °C temperature and 100% humidity, the emptying interval of water trap is as follows.
For Adult/Child
17 hours (@200 mL/min.)
26 hours (@120 mL/min.)
For Neonate