Chapter 5 Initial Settings User I/F
When the discharge process is performed, patient data such as recall and trend will be
initialized. The parameter and alarm settings will be reset according to the settings made
under [Menu>Initial Settings>User I/F>Power ON/Discharge).
The operation after the power is turned ON will be according to the setting made on [Initial
Settings] > [User I/F] > [Power ON/Discharge]. However, if the power was turned OFF for
less than 30 seconds, the setting before the power was turned OFF will remain.
If the "Main Mode" setting is other than [Backup], the following cannot be set.
Patient Classification, Alarm, Display Configuration, ECG1, ECG2 Lead, ECG1, ECG2
Size, Impedance Mode ON/OFF, CVA Detect, NIBP Auto Mode, BP Scale, SpO
2
Averaging, CO
2
Scale, EtCO
2
Peak Duration
ECG1, ECG2 Size Initialize The setting will be initialized with the currently selected mode.
Impedance
Mode ON/OFF
Initialize The setting will be initialized with the currently selected mode.
CVA Detect OFF CVA detection will be set to OFF.
NIBP Auto Mode OFF NIBP auto mode will be turned OFF.
OFF->2.5 min. If NIBP Auto Mode is OFF, 2.5 min. interval will be set.
OFF->5 min. If NIBP Auto Mode is OFF, 5 min. interval will be set.
2.5 min. NIBP auto mode will be set to 2.5 min. interval.
5 min. NIBP auto mode will be set to 5 min. interval.
BP Scale Initialize The setting will be initialized with the currently selected mode.
SpO
2
Averaging Initialize The setting will be initialized with the currently selected mode.
CO
2
Scale Initialize The setting will be initialized with the currently selected mode.
EtCO
2
Peak Duration 10 sec. EtCO
2
peak picking duration will be set to 10 sec.
Selection other than Backup
Item Setup Power ON/Discharge