9-3
Chapter 9 Maintenance Check Daily and Periodic Check
If the CO
2
gas calibration is not performed at a specified interval, CO
2
measurement
accuracy may be affected and also subsequent gas calibration may not be possible.
10 Function (When MGU-810 is used) The home display should appear, and the power
LED located at the lower right of the MGU-810
should light.
11 Function (When HP-800/HM-800/HG-800 are used) <Check Connection> message for the module
should not be displayed.
Any error message for the expansion module
should not be displayed on the home display.
12 Function
(When the DSA-82
is used)
Turn ON the power of the host monitor, and
check whether the power supply LED lights.
The power supply LED on the front side of the
adapter should light.
13 Cables Visually check all cables for any damage. No damage should be found.
14 CO
2
Calibration
(When HCP-800/
810/820 is used)
Check the date of the previous calibration.
(Refer to the following caution.)
Should be within one year.
Check the remaining time until the next
calibration.
[Menu]> [CO
2
]> [CO
2
Cal.]
Should not be 0 hrs.
15 Sampling Tube
(When MGU-800/
MGU-810 is used)
Check that it is not reused. Extra supplies should be kept.
16 Airway Adapter
(When MGU-800/
MGU-810 is used)
Check that it is not reused. Extra supplies should be kept.
17 Flow Sensor
(When MGU-810 is
used)
Check that it is not reused. Extra supplies should be kept.
18 Gas Measurement
Accuracy Check
(When MGU-800/
MGU-810 is used)
Check the date of the previous gas
measurement accuracy check.
Should be within one year.
No. Check Items Check Procedure Criteria
4. Periodic Replacement Parts, Periodic Check
01 Water Trap (When
MGU-800/MGU-
810 is used)
Check the last replaced date of reservoir. Should be within 1 month.
02 Nafion Tube
DRYLINE
Receptacle (When
MGU-800/MGU-
810 is used)
Check the last replaced date of Nafion Tube,
O
2
filter, dust filter.
Should be within 1 years.
03 Periodic Inspection Check the date of the previous periodic
inspection.
Should be within one year.