1-13
Chapter 1 Installation of the Unit System Construction
Connecting the Exhaust Tube
1Connect the exhaust tube to the MGU-800/MGU-810 series.
Connect one end of the exhaust tube to the exhaust hole of the multigas unit, and the other end to the gas
exhaust system in the hospital.
1 Exhaust Hole
2 Exhaust Tube
3 To the gas exhaust system
WARNING
To protect the hospital staffs from unnecessary anesthetic agent, it is strongly
recommended to connect the exhaust hole to the gas exhaust system in the hospital.
HR-800 Recorder Unit Connection Example 1
1Connect the main unit and the HR-800 with the unit connection cable (CJO-09SSxx).
1 Main Unit (DSC-8500 series)
2 Unit Connection Cable (CJO-09SSxx)
3 HR-800
1
2
3
1
2
3