2-4
Chapter 2 Network System Construction Wired Network System
4 Enter the Bed ID using the numeric keypad.
REFERENCE
To display the keypad for Room ID again, press the input area for the Room ID.
5 Press the [Regist] key.
The entered ID will be displayed on the upper left of the screen.
4Set the "DS-LAN Pat. ID Transmission Start Position
REFERENCE
On the DS-8500 system, patient ID of up to 20 digits can be set, but only 10 digits can be
transmitted on a DS-LAN II network.
This setup will set the starting digit from the 20 digits to be transmitted on the DS-LANII
network.On the DS-LANIII network, if [Central] is selected for the printer and printing is
started on the bedside monitor, the central monitor printer can print only up to 10
digits.This setup allows to set the starting digit of the 10 digits to be printed. 20 digits can
be transmitted on the DS-LAN III network.
1 Press the key for "DS-LAN Pat. ID Transmission Start Position".
The "DS-LAN Pat. ID Transmission Start Position" window will be displayed.
2 Enter the starting position in the range from 1 to 20.
5Set the "Synchronize Hemodynamic Data with the Central Monitor".
1 Press the key for "Synchronize Hemodynamic Data with the Central Monitor".
The dropdown list will be displayed.
2 Select [ON] or [OFF].
[ON]:5 latest hemodynamic data will be synchronized between this monitor and the central monitor.
Other hemodynamic data will be deleted.
When the hemodynamic data is edited on this monitor, the result will be also reflected on the central
monitor, and vice versa.
[OFF]:5 latest data will be transmitted to the central monitor, but the data will not be synchronized
between this monitor and the central monitor. The hemodynamic data edited on the central monitor will
be deleted. The hemodynamic data edited on this monitor will be transmitted to the central monitor.
6Set the "CO
2
(mmHg) Upper Limit of Transmission".
REFERENCE
If the CO
2
measurement unit is "mmHg", and the CO
2
value is 100mmHg or above,
whether or not to limit the value for transmission to the central monitor can be set.
1 Press the key for "CO
2
(mmHg) Upper Limit of Transmission".
The dropdown list will be displayed.
2 Select from [No limit]/[99mmHg].
[No limit]: Actual CO
2
value will be transmitted to the central monitor even if the value is 100mmHg or
above.
[99mmHg]: 99mmHg will be transmitted as the CO
2
value if the value is 100mmHg or above.