c. Thread the set screw into the threaded hole in the guide rail near the “E” or “W” inscription. Refer to
Figure 5-10. The socket of the set screw should be on the outside of the canopy.
OUTSIDE OF CANOPY
INSIDE OF CANOPY
Larger lobe
Long lip of seal
Canopy edge
NOTE: The canopy
is upside-down.
Lift assembly to upper limit of canopy holes.
Figure 5-12 Seal Cross-section
d. At the middle position of the long canopy sides, install the longer screw, lockwasher, spacer washer,
gasket, door support, and smooth nut, as shown in Figure 5-10.
e. Slide the seals on both sides, for alignment purposes. Lift the seal assembly to the upper limit of the
canopy holes. Refer to Figure 5-12.
f. Using a 10 mm open-end wrench, tighten all nuts and screws until they feel snug.
g. Remove seal and verify that the lock washers are compressed at. If necessary, tighten with 10 mm
open-end wrench and 2.5 mm hex key.
CAUTION:
Do not over-tighten the hardware.
h. Install and remove seals to check for easy installation.
80 6600-0343-000 104 © 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 5: Repair Procedures

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed and is the answer not in the manual?

GE Giraffe OmniBed Specifications

General IconGeneral
Temperature ControlYes
Humidity ControlYes
Servo Humidity ControlYes
Oxygen DeliveryYes
ScaleIntegrated
Trendelenburg/Reverse Trendelenburg±12°
Rotating MattressNo
X-ray CompatibilityYes
MobilityYes, with casters
CategoryIncubator
Humidity Range30% to 95%
Oxygen HoodYes
Access PortsMultiple
ModesAir Mode, Skin Mode
Power Requirements100-240V AC, 50/60 Hz

Related product manuals