5.9 Lower Unit Repairs
5.9.1 Removing the Chassis Cover with the Storage Drawer in Place
Refer to “Figure 5-54 Chassis Bottom Cover.
NOTE: It is not necessary to remove the storage drawer before removing the chassis cover, but you may wish
to remove it to provide easier access to the chassis cover.
1. Remove the humidier reservoir.
2. Slide the drawer all the way over in one direction, and use a 2.5 mm hex key to remove the 3 chassis cover
screws on one side.
3. Slide the drawer to the other side and remove the 3 remaining screws from the cover.
4. Remove the bottom 2 screws (2mm hex key) that secure the right end of the humidier wire raceway cover.
5. Flex the raceway cover and carefully slide the cover panel forward until it drops o the back drawer slide.
6. Rotate the cover slightly and push it out the right side of the unit.
7. Remove the ground wire.