e. If your canopy has six holes across the north end, install parts for the middle four holes as shown in
Figure 5-13 and install parts for the remaining two holes as shown in Figure 5-15. The parts for these
two holes serve to ll these two holes.
Smooth nut
Canopy
Splitlock washer
M4x12 buttonhead
stainless steel screw
Spacer washer
OUTSIDE OF
CANOPY
INSIDE OF
CANOPY
Flat washer
Figure 5-15 Hole Fill Assembly (2x) for Canopies with Six Holes across North End
f. Snap the north seal vertically between the north spacer and canopy for a t-check. Make sure that the
red/orange indicator is not visible from inside the canopy.
g. Using a 10 mm open-end wrench, tighten all smooth nuts.
CAUTION:
Do not over-tighten the hardware.
h. Remove the north seal.
82 6600-0343-000 104 © 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 5: Repair Procedures

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed and is the answer not in the manual?

GE Giraffe OmniBed Specifications

General IconGeneral
Temperature ControlYes
Humidity ControlYes
Servo Humidity ControlYes
Oxygen DeliveryYes
ScaleIntegrated
Trendelenburg/Reverse Trendelenburg±12°
Rotating MattressNo
X-ray CompatibilityYes
MobilityYes, with casters
CategoryIncubator
Humidity Range30% to 95%
Oxygen HoodYes
Access PortsMultiple
ModesAir Mode, Skin Mode
Power Requirements100-240V AC, 50/60 Hz

Related product manuals