2.6.2.2 O Center Weight Check
1. Enter the service screen by holding the override key (>37) while powering up. Set the scale resolution to 5
grams (if option is available).
2. Shut o the unit and power it back up to enter normal weighing mode.
3. Place a known test weight 10 cm (4”) from the center of the mattress in 4 positions 90 degrees from each
other (for example- toward each corner of the mattress) and check the reading at each position. Resulting
weight readings should be within +/- 10 g of previously obtained center weight reading.
2.7 Other Accessory Checks
1. Check that all accessories are securely mounted and out of the path of the canopy.
2. Check the operation of any accessories with reference to their appropriate operation manuals.
3. Set up any required suction or gas supply systems. Check them for leaks as described in their respective
operation manuals.