GE Giraffe OmniBed - Baby Mode Disabled Alarm; Baby Probe 1 Failure or Baby Probe 2 Failure Alarm; Bad Membrane Switch Alarm

GE Giraffe OmniBed
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.3.4 “Baby Mode Disabled” Alarm
Cause or Circumstances Failure Action
Baby mode has been disabled
in the service screen and
baby mode button is pressed.
None
Press alarm silence button to deactivate the alarm.
If you wish to use the baby mode, re-enable it in service
screen.
4.3.5 “Baby Probe 1 Failure” or “Baby Probe 2 Failure” Alarm
Cause or Circumstances Failure Action
Probe thermistors show
dierence > 0.5ºC.
Defective patient probe Replace patient skin probe.
Defective patient probe jack
wiring
Plug patient probe into probe jack 1. In service mode,
note the dierence between readings from P11 and
P12. Plug the same probe into probe jack 2 and note the
dierence between readings from P21 and P22. If they
are greater there is resistance in the probe jack harness.
Replace the harness.
4.3.6 “Bad Membrane Switch” Alarm
Cause or Circumstances Failure Action
All
Defective membrane switch
panel
Test switches in service mode. Replace membrane
switch panel.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved. 6600-0343-000 104 35
Chapter 4: Troubleshooting