5.5.5 Replacement of the Left Upright Heater Wire Harness/Cable Carrier Assembly
The heater wire harness/cable carrier assembly (6600-0710-700) includes both the wire harness and the cable
carrier. The end links of the cable carrier are riveted to the lift rail and the cable carrier channel (2 pop rivets at
each end). The easiest way to replace the harness is to reuse these end links of the cable carrier. If the end link
of the cable carrier is damaged, there are two options:
Replace the entire left upright assembly (66600-0288-851). (Refer to section 5.4.2)
Drill out the rivets that secure the broken link to the lift rail or Cable carrier channel. Remove the end link
from the new assembly and rivet it in place. Then follow the instructions below. Two rivets are required for
each end of the cable.
6600-1160-400 Pop Rivet Stainless Steel, Body Diameter .125”, body length .313”
The 3-pin MAT-N-LOK connector for the radiant heater connections must be removed from the old harness
using a pin extractor. If a pin extractor is not available, order a new connector when you order the harness.
Four cable ties are also required.
6600-1097-603 3 pin Universal MAT-N-LOK connector
6600-0384-400 Cable Tie
5.5.5.1 Removing and Disassembling the Rail
Refer to “Figure 5-33 Left Rail Springs and Spools”.
1. Remove the left upright following section 5.4.2.
2. Remove the inner rail assembly following section 5.4.3.1.
3. Remove the 3 buoyancy spring spools from the cable carrier channel. Save the spring spools and plastic
washers for reinstallation.
5.5.5.2 Removing the Old Harness Assembly from the Left Rail
For instructions on removing the harness from the cable carrier channel, refer to section 5.5.5.4.
NOTE: Before removing the old harness make note of the wire routing, cable tie orientation and placement,
and bushing location. This will assist in installing the replacement harness.
106 6600-0343-000 104 © 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 5: Repair Procedures

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed and is the answer not in the manual?

GE Giraffe OmniBed Specifications

General IconGeneral
BrandGE
ModelGiraffe OmniBed
CategoryMedical Equipment
LanguageEnglish

Related product manuals