Appendix B: Specications
B.1 Power Requirements
10.5 A @100v ~, 50/60 Hz
9 A @115v ~, 50/60 Hz
4.5 A @ 220v ~, 50/60 Hz
4.5 A @ 230v ~, 50/60 Hz
4.5 A @ 240v ~, 50/60 Hz
Inrush for 1/2 cycle current < 80 A
Radiant Heater Element = 450-480 Watts
B.2 Accessory Outlets
2 A @100v ~, 50/60 Hz
2 A @115v ~, 50/60 Hz
1 A @ 220v ~, 50/60 Hz
1 A @ 230v ~, 50/60 Hz
1 A @ 240v ~, 50/60 Hz
B.3 Standards
Designed to meet requirements of
IEC 60601-2-19 1990 with amendment 1 1996
IEC 60601-2-21 1994 with amendment 1 1996
IEC 60601-1 1998 with amendment 1 1991 and amendment 2 1995
IEC 60601-1-2 2007
21 CFR CH-1, Section 1020.30 (N)
UL 60601-1 2003
CAN/CSA-C22.2 No. 601.1-M90 1990 with supplement 1 1994 and amendment 2 1998
ISO 19054 2005 with an exception claim to clause 5.2.7.2. Slight deformation of rail occurs after application
of 500N to wide surface of DIN rail.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved. 6600-0343-000 104 235

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Giraffe OmniBed and is the answer not in the manual?

GE Giraffe OmniBed Specifications

General IconGeneral
Temperature ControlYes
Humidity ControlYes
Servo Humidity ControlYes
Oxygen DeliveryYes
ScaleIntegrated
Trendelenburg/Reverse Trendelenburg±12°
Rotating MattressNo
X-ray CompatibilityYes
MobilityYes, with casters
CategoryIncubator
Humidity Range30% to 95%
Oxygen HoodYes
Access PortsMultiple
ModesAir Mode, Skin Mode
Power Requirements100-240V AC, 50/60 Hz