GE HealthCare CONFIDENTIAL
Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 7 Diagnostics/Troubleshooting 7 - 19
7-3-5 Home (continued)
System Status
Use System Status to view status information on the LOGIQ Totus. Specifically, do the following:
• View the hardware configuration file
• View temperatures for FRUs and a graphical representation of the monitored temperatures
• View statuses of the batteries
• View voltages for FRUs.
• View issues with the boot process
• View alert logs
• View logs that occurred before and after an error
• View USB devices connected to the LOGIQ Totus
• View the amount of time the LOGIQ Totus has been running
• View the amount of time since the LOGIQ Totus has last been rebooted
The information on System Status is available to Class C and Class M licenses.
To access System Status, navigate to Utility (second page) > Service > Home.
Figure 7-10 System Status