Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
7 - 20 Section 7-3 - Service Desktop
7-3-5 Home (continued)
This table shows all the elements available on System Status with descriptions.
Table 7-4 System Status
Element Description
Hardware Configuration Displays the contents of the hardware configuration file.
Temperature
Displays the temperature monitoring of the LOGIQ Totus as normal, warning, or
error using the standard colors and icons. Includes temperature out-of-spec events
including an overall count of event warnings and errors for the last 30 days or the
last status checkpoint whichever is sooner. Temperature displays temperature a
graph for the configured monitoring points.
Battery Status Displays the battery monitoring of the LOGIQ Totus.
Voltage
Displays the voltage monitoring of the LOGIQ Totus as normal, warning or error
using the standard colors and icons. Includes voltage out-of-spec events including
an overall count of the event warnings and errors for the last 30 days or last status
checkpoint whichever is sooner.
Bootup Issues
Based on the latest occurrence, displays all abnormal boots of the LOGIQ Totus for
either the past 30 days or the latest check point.
Alert Logs
Based on the latest occurrence, displays all alerts generated by the LOGIQ Totus
for either the past 30 days or the latest check point.
Log Analyzer Displays a list of errors and associated log files for troubleshooting.
USB Devices Displays the USB connected devices and their status.
System On Time Displays the time the LOGIQ Totus was turned on.
Time Since Last Reboot Displays the time the LOGIQ Totus was last rebooted.