GE HealthCare CONFIDENTIAL
Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 7 Diagnostics/Troubleshooting 7 - 35
7-3-7 Utilities (continued)
This table shows all the elements available on Network Capture with descriptions.
Table 7-12 Network Capture
Element Description
Network Capture Parameters
Maximum Size
Allowed size of the generated log file. Valid value are:
• 256MB
• 512MB
• 1024MB
Devices
DICOM-configured devices for which you want to capture information. If no
additional devices are configured, only All will be available.
Start Network Capture
Select to start the process. This causes the network capture to start, enables the
Stop button, and updates the Network Capture Status pane and changes the
Status to Running.
Stop Network Capture Select to stop the process.
Network Capture Status
Displays information about the status of the network capture. The language setting
for this information is set in Windows and not through the Service desktop or
LOGIQ Totus application software.
Displays the current status of the network capture. Valid values are:
• Not Running
• Running